LUTRON YK-2005WA PH/ORP/DO/ TDS/ CONDUCTIVITY
LUTRON YK-2005WA PH/ORP/DO/ TDS/ CONDUCTIVITY
LUTRON YK-2005WA adalah alat ukur kualitas air multifungsi (multi-parameter water analysis meter) portabel yang menggabungkan lima fungsi utama, yaitu pengukuran pH, ORP, Dissolved Oxygen (DO), Konduktivitas (CD), dan TDS dalam satu perangkat. Alat ini diproduksi oleh Lutron Taiwan dan dirancang untuk kebutuhan laboratorium, pengolahan air, serta riset lingkungan.
Lutron YK-2005WA PH/ORP, DO, CD/TDS Meter has an option of automatic temperature compensation and a real-time data logger that records 16,000 data (year, month, date, hour, minute, second).
Lutron YK-2005WA PH/ORP, DO, CD/TDS Meter is a one meter for multi purpose operation which measures PH/ORP, DO & CD/TDS whose ranges are pH : 0 to 14.00 pH, ORP : 1999 mV, Conductivity : 200 uS/2 mS/20 mS/200 mS, Dissolved oxygen : 0 to 20.0 mg/L.
Applications: Used for oxidation/reduction measurement in plating baths, wastewater monitoring, and other applications such as Aquarium, Beverage, Fish Hatcheries, Food Processing, Laboratory, and Quality Control.
Features Lutron YK-2005WA pH/ORP, DO, CD/TDS Meter
- One meter for multi purpose operation : PH/ORP, DO, CD/TDS METER
- pH : 0 to 14.00 pH, ORP : ± 1999 mV
- Conductivity : 200 uS/2 mS/20 mS/200 mS
- Dissolved oxygen : 0 to 20.0 mg/L
- ATC (automatic temperature compensation)
- Real time data logger ( record year, month, date, hour, minute, second ), 16,000 data logger
- Max., Min., Data hold
- RS-232 computer interface
- Optional electrode : pH, ORP, CD/TDS, DO, ATC probe
- Jumbo LCD display
- DC 1.5V ( UM-3, AA ) x 4 PCs or
DC 9V adapter in. - Size : 203 x 76 x 38 mm
Specifications Lutron YK-2005WA pH/ORP, DO, CD/TDS Meter
General Specifications |
|
| Circuit | Custom one-chip of microprocessor LSI circuit. |
| Display | LCD size : 58 mm x 34 mm. |
| Measurement | PH/ORP, Conductivity/TDS(Total Dissolved Solids) & Dissolved Oxygen |
| Sampling Time of Data Logger | Manual : Push the data logger Button once will save data one time. @Set the sampling time to 0 second Auto : 1 sec to 8 hour 59 min. 59 sec. |
| Data Hold | Freeze the display reading. |
| Memory Recall | Maximum & Minimum value. |
| Power off | Auto shut off saves battery life or manual off by push Button. @Can default auto power or manual power off @When default auto power function, power will off automatically after 10 MIN, if no Button be pressed. |
| Accessories Included | * Instruction manual * Hard Case |
| Optional Accessories | PH electrode PE-03, PE06HDA ATC Temp Probe TP-07 pH 7 buffer solution PH-07 pH 4 buffer solution PH-04 Conductivity probe CDPB-03 1.413 mS Conductivity Standard Solution CD-14 ORP Probe ORP14 Oxygen probe OXPB-11 + OXEl-03. USB cable USB-01 Data Acquisition software, SW-U801-WIN. Data Logger software, SW-DL2005. |
Electrical Specifications |
|
| PH Electrode | Optional, Any PH electrode with BNC connector. |
| Measurement | PH : 0 to 14 PH mV : -1999 mV to 1999 mV |
| Input Impedance | 10^12 ohm |
|
Temperature Compensation
for pH measurement |
Manual : 0 to 100 ℃ , be adjusted by push button on front panel. |
| Automatic ( ATC ) : With the optional temperature probe ( TP-07 ) 0 to 65 ℃. | |
| pH Calibration | PH7, PH4, and PH10, 3 points calibration ensure the best linearity and accuracy |
| Optional probe and accessories | * PH electrode………….PE-03, PE-11, PE-01, PE06HD PE-04HD, PE-05T, PE-03K7 * ATC ( automatic temperature ) TP-07 * pH 7 buffer solution………………… PH-07 * pH 4 buffer solution…………………. PH-04 * ORP electrode………………………… ORP-14 |
2-2 Electrical Specifications |
|||
| Measurement | Range | Resolution | Accuracy |
| PH | 0 to 14 PH | 0.01 PH | ± (0.02 PH + 2 d) |
| mV | 0 to 1999 mV | 1 mV | ± (0.5% + 2 d) |
| B. Conductivity | |
| Conductivity probe | Optional, Carbon rod electrode for long life. |
| Function | * Conductivity ( uS, mS ) * TDS ( Total Dissolved Solids, PPM ) * Temperature ( , ) |
| Probe Dimension | Round, 22 mm Dia. x 120 mm length. |
| Optional probe and accessories | * Conductivity probe……………………CDPB-03 * 1.413 mS Conductivity Standard Solution……..CD-14 |
| 1. Conductivity ( uS, mS ) | |||
| Range | Measurement | Resolution | Accuracy |
| 200 uS | 0 to 200.0 uS | 0.1 uS |
± (2% F.S.+1d)
* F.S. – Full scale |
| 2 mS | 0.2 to 2.000 mS | 0.001 mS | |
| 20 mS | 2 to 20.00 mS | 0.01 mS | |
| 200 mS | 20 to 200.0 mS | 0.1 mS | |
| 2. TDS ( Total Dissolved Solids ) | |||
| Range | Measurement | Resolution | Accuracy |
| 200 PPM | 0 to 132 PPM | 0.1 PPM |
± (2% F.S.+1d)
* F.S. – Full scale |
| 2,000 PPM | 132 to 1,320 PPM | 1 PPM | |
| 20,000 PPM | 1,320 to 13,200 PPM | 10 PPM | |
| 200,000 PPM | 13,200 to 132,000 PPM | 100 PPM | |
| 3. Temperature | |||
| Range | Measurement | Resolution | Accuracy |
| ℃ | 0 ℃ to 60 ℃ | 0.1 ℃ | ± 0.8 ℃ |
| ℉ | 32 ℉ to 140 ℉ | 0.1 ℉ | ± 1.5 ℉ |
| C. Dissolved oxygen | |||
| Oxygen Probe | Optional, The polarographic type oxygen probe with | ||
|
Measurement & Range
|
Dissolved Oxygen | 0 to 20.0 mg/L ( liter ). | |
| Oxygen in Air | 0 to 100.0 %. | ||
| Temperature | 0 to 50 . ℃ | ||
|
Resolution
|
Dissolved Oxygen | 0.1 mg/L. | |
| Oxygen in Air | 0.1 % O2 . | ||
| Temperature | 0.1 . |
|
|
|
Accuracy
|
Dissolved Oxygen | ± 0.4 mg/L. | |
| Oxygen in Air | ± 0.7% O2. |
|
|
| Temperature | ± 0.8 ℃ /1.5 ℉. | ||
|
Probe Compensation & Adj.
|
Temperature | 0 to 50 ℃, Automatic |
|
| Salt | 0 to 39 % Salt | ||
| Height ( M. T.) | 0 to 8900 meter |
|
|
| Probe Weigh | 335 g/0.74 LB ( batteries & probe included ) | ||
| Probe Size |
190 mm x 28 mm Dia. ( 7.5″ x 1.1″ Dia. )
|
||
| Optional Accessories |
* Oxygen probe……………………………..OXPB-11
* Spare Probe head with Diaphragm set ….. OXHD-04 * Probe-filling Electrolyte……………….. OXEL-03 |
||

Mohon Tanyakan Stok Terlebih Dahulu Sebelum Order.